The Journal of Toxicological Sciences
Online ISSN : 1880-3989
Print ISSN : 0388-1350
ISSN-L : 0388-1350
Letter
Design and performance of a system for blood collection of rats under whole-body inhalation exposure
Makoto TakeMakoto OhnishiKasuke NaganoSeigo YamamotoShoji Fukushima
Author information
JOURNAL FREE ACCESS

2009 Volume 34 Issue 2 Pages 221-226

Details
Abstract

In order to obtain basic risk assessment data on human health exposure to volatile organic compound (VOC) vapor by inhalation, a whole-body inhalation exposure system which allows blood collection during the exposure period was designed. The system was tested using chloroform as a model VOC. Chloroform vapor, sampled from the supply-header, animal-chambers and exhaust-header, remained constant in this system with variations in its concentration being less than 2%; flow rate of the vapor through the system was also constant. Rats were exposed to chloroform vapor and blood collected from the tail during exposure to the chloroform vapor. The chloroform concentration in the blood increased during the initial 60 min of exposure, and afterwards its concentration remained at about 2 µg/ml from 60 to 360 min. In conclusion, our design allows blood to be collected from individual rats during exposure by inhalation to test VOCs and changes in the blood concentration of the VOC during exposure to be assessed.

Content from these authors
© 2009 The Japanese Society of Toxicology
Previous article Next article
feedback
Top