可視化情報学会誌
Online ISSN : 1884-037X
Print ISSN : 0916-4731
ISSN-L : 0916-4731
C202 MERAモノピボット遠心血液ポンプの流れの可視化解析
木暮 尚登西田 正浩丸山 修小阪 亮山根 隆志河村 洋山本 好宏桑名 克之山海 嘉之筒井 達夫
著者情報
ジャーナル フリー

2006 年 26 巻 Supplement1 号 p. 183-184

詳細
抄録
A circulatory assist pump, which can be used for around two weeks after an operation, is under development by MERA. For R&D of a centrifugal blood pump it is important to prevent the thrombus formation caused by low shear stress and the hemolysis caused by high shear stress. Flow visualization analysis was performed around the pivot to prevent thrombus formation. We analyzed horizontal and vertical section around the pivot and enough washout to prevent thrombus formation was verified. In addition, a slight stagnant region was found in the outlet section and a relatively weak jet in the impeller tip region. Hemolysis tests were conducted also under the conditions of 100mmHg-5L/min and 200mmHg-5L/min. Hemolysis index, NIH, was 150% of BPX-80 under the condition of 100mmHg-5L/min, and 80% of BPX-80 under the condition of 200mmHg-5L/min. The hemolysis rate improves for high pressure conditions.
著者関連情報
© 社団法人 可視化情報学会
前の記事 次の記事
feedback
Top