Journal of the Vacuum Society of Japan
Online ISSN : 1882-4749
Print ISSN : 1882-2398
ISSN-L : 1882-2398
研究論文
Attempts to Improve the Sensitivity and the Energy Resolution of an Analyzer for Auger Photoelectron Coincidence Spectroscopy and Electron Ion Coincidence Spectroscopy
Sadanori ARAERui KANEMURAKenta HIRAGAYosuke OGASHIWAKohtaro YANASENoritsugu KANAYAMAShinya OHNOTakuhiro KAKIUCHIKazuhiko MASEKoji K. OKUDAIRAMakoto OKUSAWAMasatoshi TANAKA
著者情報
ジャーナル フリー

2013 年 56 巻 12 号 p. 507-510

詳細
抄録
  We constructed two types of coincidence analyzers those can be used for Auger photoelectron coincidence spectroscopy and electron ion coincidence spectroscopy. The first one is designed for high electron energy resolution measurements, and consists of a double-pass coaxially symmetric mirror electron energy analyzer (DP-ASMA), a double-pass cylindrical mirror electron energy analyzer (DP-CMA), a time-of-flight ion mass spectrometer (TOF-MS), a magnetic shield, a conflat flange, an xyz stage, and a tilt adjustment mechanism. The second one is dedicated for high sensitivity measurements, and involves an ASMA with a large pinhole, DP-CMA with a large pinhole, and a TOF-MS.
著者関連情報
© 2013 The Vacuum Society of Japan
前の記事 次の記事
feedback