日本機械学会論文集 C編
Online ISSN : 1884-8354
Print ISSN : 0387-5024
MRI装置用勾配コイルドラムのインテリジェント化
裘 進浩谷 順二大友 規矩雄
著者情報
ジャーナル フリー

1995 年 61 巻 586 号 p. 2412-2417

詳細
抄録

High-level noise induced in MRI equipment has been a main obstacle for their wide application in medical diagnosis. The noise is caused by the vibration of the cylindrical shell used to support the gradient magnetic field coils in MRI. In this study, a reduced-size model of the coil drum is constructed using polymer films and piezoelectric films. The piezoelectric films are divided into several patches and two patches of them are used as integrated bending actuators in the control of the shell vibration so that no external actuators are needed. The disturbance cancellation method is used since the shell is excited by predetermined forces. The relationships between the size, position and output magnitude of the actuators and control effect are also discussed. Experiment and simulation were carried out for the cases where the shell was excited at resonance frequencies of three main modes. The results show that the disturbance cancellation method in combination with the piezoelectric film actuators can successfully suppress the forced vibration of the shell and that the minimum amplitude of controlled vibration varies with the resonance mode.

著者関連情報
© 社団法人日本機械学会
前の記事 次の記事
feedback
Top