日本機械学会論文集 C編
Online ISSN : 1884-8354
Print ISSN : 0387-5024
高安全水圧駆動方式能動カテーテルの研究(機械力学,計測,自動制御)
生田 幸士市川 尋信鈴木 克也松田 喜勝矢島 大輔
著者情報
ジャーナル フリー

2010 年 76 巻 766 号 p. 1553-1559

詳細
抄録

We aimed to develop an active catheter that could be used safely in human blood vessels. The various types of active catheter were proposed to realize bending degrees of freedom and to select courses at a branch point of a blood vessel easily. Since conventional ones were developed with electric actuator like shape memory alloy or conductive polymer and so on, they might damage the patient by its accidental electric leakage. Now, we proposed a new type micro valve and incorporated it into a water pressure-driven system with multi-degree of freedom, to produce an active catheter that was compact and required no electrical power for operation. We actually achieved to develop this valve by applying hybrid micro stereolithography and produced the active catheter with two bending segments successfully. In addition, we experimentally verified the good operational performance of a prototype.

著者関連情報
© 2010 社団法人日本機械学会
前の記事 次の記事
feedback
Top