Patient-Specific Femoral Guide with 3-Dimensional Support Software in Anterolateral Supine Approach of Total Hip Arthroplasty: A Single-Arm Trial
DAIHEI KIDAHIROYA HASHIMOTONORIKO ITOYUKARI KITOKOUICHI MORIYOSUKE HATTORINOBUNORI TAKAHASHIMASAAKI MATSUBARA
Author information
JOURNAL FREE ACCESS

2021 Volume 68 Issue 1 Pages 1-8

Details
Abstract

Summary: Currently, there is no surgical assistance system that can perform a three-dimensional (3D) planned total hip arthroplasty (THA) by methods other than surgical assistance navigation or robots. However, they are expensive, cumbersome, and subject to additional invasiveness, so there is a need for a simpler and less expensive 3D surgical support system. In this study, THA was performed using the anterolateral approach (Watson-Jones) in the supine position in 23 subjects to examine the efficacy and safety of a patient-specific femoral guide linked to 3D surgery support software. In 48% of the subjects, the difference in anterior torsion angle from the preoperative plan was within ±5 degrees, while in 83% of the subjects, the difference was within ±10 degrees. The 95% confidence interval (4.61-8.70) of the absolute difference did not fall below the pre-defined threshold of 7.2 degrees (p = 0.293). No adverse events were observed other than 2 cases (8.7%) of hemorrhage that required a blood transfusion. We confirmed the efficacy and safety of the patient-specific femoral guide in anterolateral supine approach THA.