リソグラフィー用KrFエキシマレーザーの出力スペクトルの高精度評価
若林 理越智 秀臣溝口 計内野 喜一郎
著者情報
ジャーナル フリー

2003 年 31 巻 7 号 p. 482-488

詳細
抄録
A measuring method to evaluate the spectral profile of a KrF excimer laser used in lithography has been developed. In this method, the parameter of the spectral purity is defined by the spectral width of the laser (Δλ95%E), which contains 95 % of its total energy. During the measurements, the instrumental function of a high-resolution spectrometer was precisely estimated with a single-mode laser. The deconvolution technique was applied to evaluate the spectral width of the laser from the observed spectrum by using the instrumental function and the influence on the deconvolved profiles, due to noises included in the observed spectra, was examined quantitatively. In order to keep the required measurement accuracy within ± 5 % of the value Δλ95%E, the necessary product value of the maximum signal count-number and the signal averaging number was found to be more than 3.65×106 in the present system.