レーザー研究
Online ISSN : 1349-6603
Print ISSN : 0387-0200
ISSN-L : 0387-0200
レーザーオリジナル
レーザープラズマ光源を用いた卓上型軟X線顕微鏡イメージング
中島 一久
著者情報
ジャーナル フリー

2004 年 32 巻 12 号 p. 786-792

詳細
抄録
We developed laser-produced plasma X-ray sources using femtosecond laser pulses at 10 Hz repetition rate in a table-top size to investigate basic mechanism of X-ray emission from laser-matter interactions and its application to an X-ray microscope. In a soft X-ray region, laser-plasma X-ray emission from a solid target achieved an intense flux of photons of the order of 1011 photons/rad per pulse with duration of a few hundreds ps, which is intense enough to make a clear imaging in a short time exposure. As an application of laser-produced plasma X-ray source, we developed soft X-ray imaging microscope operating at the wavelength of 13.9 nm. The microscope consists of a cylindrically ellipsoidal condenser mirror and a Schwarzschild objective mirror with highly-reflective multilayer. We report results of performance tests and biological cell observations of the soft X-ray imaging microscope with a compact laser-produced plasma X-ray source.
著者関連情報
© 2004 一般社団法人 レーザー学会
前の記事 次の記事
feedback
Top