Abstract
Dynamic response of an electro-striction assembly for laser frequency control has been measured by a simple method and been theoretically analyzed. The resonantphenomenon of the electro-striction assembly occurs at the rather lower frequency than its natural frequency of the electro-striction element. Its resonant frequency was given as a function of a laser mirror mass and a compliance of silicon glue which used to the assemble of parts such as electro-striction element, laser mirror and so on. The experimental results were in fairly well agreement with theoretical ones within a measurement error.