Mixture of Caesium Iodide, Sodium Iodide, and Glycerol as a Calibrant for Routine Fast Atom Bombardment Mass Analysis
Takemichi NakamuraHidemi NagakiTakeshi Kinoshita
著者情報
ジャーナル フリー

1988 年 36 巻 2 号 p. 81-86

詳細
抄録
A mixture of caesium iodide, sodium iodide, and glycerol was found to be a convenient calibrant for routine fast atom bombardment mass analysis. The mixture can be used for both positive and negative ion modes, and at a mass range around m/z 10 to 2,500 with reasonable mass accuracy.