報文
負イオン化学イオン化質量分析法における低質量域校正用標準物質の検討
柴田 明宏飯田 芳男八木 宏憲
著者情報
ジャーナル フリー

1991 年 39 巻 2 号 p. 79-91

詳細
抄録

The mass calibration standard materials in the low mass region for negative ion chemical ionization (NCI) mode in a magnetic sector mass spectrometer were sought, because perfluorokerosine (PFK) and perfluorotributylamine (PFTBA) used in EI mode scarcely produce negative ions below m/z 200. The 45 fluorine-containing compounds (5 alkanes, 6 ketones, 5 esters, 5 ethers, and 3 acid anhydrides, etc.) were tested, and the 10 : 1 : 1 mixture of 3 esters [CF3COOCH3, C2F5COOC2H5, C3F7COOC2H5] and the 1 : 1 : 1 mixture of 3 acid anhydrides [(CF3CO)2O, (C2F5CO)2O, (C3F7CO)2O] were found to be advantageous. In their NCI mass spectra, ions were observed at m/z 69, 85, 113, 128, 144, 178, 194, and m/z 113, 128, 144, 163, 178, and 194, respectively, with the relative ion intensities of several to 100%. The standard materials for mass calibration and for selected ion monitoring were also purposed.

著者関連情報
© 1991 日本質量分析学会