Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
報文
Prediction of Measurement Precision Based on FUMI Theory for Quantitative Mass Spectrometry with Electron Ionization
Mitsuo TAKAYAMAYuzuru HAYASHI
著者情報
ジャーナル フリー

2000 年 48 巻 4 号 p. 248-253

詳細
抄録
The precision or relative standard deviation (RSD) of measurements in a sector type mass spectrometer equipped with an electron ionization system is examined experimentally and theoretically. The observed RSD is obtained by the usual replication and the theoretical RSD is predicted from the probabilistic properties of the baseline noise by the uncertainty theory called the FUMI theory (Function of Mutual Information). By comparing the experimental and theoretical RSD, this paper identifies two kinds of error sources: (1) for small peaks, the major cause of the measurement uncertainty is the baseline noise originating mainly from the detection unit; (2) for large peaks, the measurement error is subject to ion/particles processes such as the fluctuation of gaseous sample concentration, electron ionization efficiency, and ion beam flux and there is a lower limit of the measurement RSD (≈3%). In case 1, the measurement RSD can successfully be predicted by the theory. In case 2, the observed RSD is even larger than the theoretical RSD, and the repeated measurements are necessary to estimate the measurement uncertainty. The peak corresponding to the limit of detection (RSD=33%) is demonstrated.
著者関連情報
© 2000 by The Mass Spectrometry Society of Japan
前の記事 次の記事
feedback