Alkylated Hydroxychalcone: A Novel Matrix for Peptide Analysis by Matrix-Assisted Laser Desorption Ionization Mass Spectrometry
Manaho Yamaguchi Yuko FukuyamaShunsuke Izumi
著者情報
キーワード: MALDI-MS, matrix, hydrophobic peptides
ジャーナル オープンアクセス HTML
電子付録

2025 年 14 巻 1 号 p. A0170

詳細
抄録

In matrix-assisted laser desorption ionization mass spectrometry, a suitable matrix is often selected for the analyte. Herein, we first developed a novel matrix, alkylated hydroxychalcone (AHC), which has properties similar to alkylated trihydroxyacetophenone (ATHAP) (Anal. Chem., 85: 9444–9448, 2013) developed as a matrix for hydrophobic peptides. However, the sample-to-sample reproducibility was low because of the poor crystallinity of AHC. The crystalline morphology of AHC changed when AHC/2,5-dihydroxybenzoic acid (DHB) was used as a binary matrix. As a result, the use of AHC/DHB improved sample-to-sample reproducibility and increased sensitivity for hydrophobic peptides. Mass imaging indicated that these results were due to an increased number of sweet spots wherein the analytes were detected as ion peaks, in a matrix/analyte crystal spot.

著者関連情報
© 2025 Manaho Yamaguchi, Yuko Fukuyama, and Shunsuke Izumi

This article is licensed under a Creative Commons [Attribution-NonCommercial 4.0 International] license.
https://creativecommons.org/licenses/by-nc/4.0/