Reduced Sampling Size with Nanopipette for Tapping-Mode Scanning Probe Electrospray Ionization Mass Spectrometry Imaging
Tsuyoshi KohigashiYoichi OtsukaRyo ShimazuTakuya MatsumotoFutoshi IwataHideya KawasakiRyuichi Arakawa
Author information
JOURNALS FREE ACCESS FULL-TEXT HTML

Volume 5 (2016) Issue 2 Pages S0054

Details
Full Text-HTML Download PDF (4126K) Contact us
Abstract

Mass spectrometry imaging (MSI) with ambient sampling and ionization can rapidly and easily capture the distribution of chemical components in a solid sample. Because the spatial resolution of MSI is limited by the size of the sampling area, reducing sampling size is an important goal for high resolution MSI. Here, we report the first use of a nanopipette for sampling and ionization by tapping-mode scanning probe electrospray ionization (t-SPESI). The spot size of the sampling area of a dye molecular film on a glass substrate was decreased to 6 μm on average by using a nanopipette. On the other hand, ionization efficiency increased with decreasing solvent flow rate. Our results indicate the compatibility between a reduced sampling area and the ionization efficiency using a nanopipette. MSI of micropatterns of ink on a glass and a polymer substrate were also demonstrated.

Information related to the author
© 2016 Tsuyoshi Kohigashi, Yoichi Otsuka, Ryo Shimazu, Takuya Matsumoto, Futoshi Iwata, Hideya Kawasaki, and Ryuichi Arakawa. This is an open access article distributed under the terms of Creative Commons Attribution License, which permits use, distribution, and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
Previous article Next article

Recently visited articles
feedback
Top