Influence of Solvent Composition and Surface Tension on the Signal Intensity of Amino Acids in Electrospray Ionization Mass Spectrometry
Ami Kageyama (Kaneshima)Akira MotoyamaMitsuo Takayama
著者情報
ジャーナル オープンアクセス HTML

2019 年 8 巻 1 号 p. A0077

詳細
抄録

The influence of solvent composition and surface tension on the signal intensity of deprotonated molecules [M−H] in electrospray ionization mass spectrometry (ESI MS) was evaluated using alanine (Ala), threonine (Thr) and phenylalanine (Phe), which have differing levels of hydrophobicity. The surface tension of the ESI solution was varied by changing the ratio of the organic solvents methanol (MeOH) and acetonitrile (MeCN) in water (H2O). In ESI MS, the signal intensity of all the amino acids was increased with decreasing surface tension for the two solutions, H2O/MeOH and H2O/MeCN. The use of H2O/MeCN was more favorable for achieving a strong signal for the analytes compared to H2O/MeOH. The smaller vaporization enthalpy of MeCN compared to MeOH was proposed as one of the most plausible explanation for this. The order of the signal intensity of amino acids was Phe>Thr>Ala, the same order as their hydrophobicity. It can be practically concluded that the use of solutions with lower surface tensions and lower vaporization enthalpies would result in higher signal intensities in ESI MS.

著者関連情報
© 2019 Ami Kageyama (Kaneshima), Akira Motoyama, and Mitsuo Takayama. This is an open access article distributed under the terms of Creative Commons Attribution License, which permits use, distribution, and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
前の記事
feedback
Top