Mass Spectrometry
Online ISSN : 2186-5116
Print ISSN : 2187-137X
ISSN-L : 2186-5116
Review
High-Temperature Liquid Chromatography and the Hyphenation with Mass Spectrometry Using High-Pressure Electrospray Ionization
Lee Chuin Chen
Author information
JOURNALS FREE ACCESS FULL-TEXT HTML

2019 Volume 8 Issue 2 Pages S0079

Details
Abstract

Increasing the operating temperature of the liquid chromatography (LC) column has the same effect as reducing the diameter of the packing particles on minimizing the contribution of C-term in the van Deemter equation, flattening the curve of plate height vs. linear velocity in the high-speed region, thus allowing a fast LC analysis without the loss of plate count. While the use of smaller particles requires a higher pumping pressure, operating the column at higher temperature reduces the pressure due to lower liquid viscosity. At present, the adoption of high-temperature LC lags behind the ultra-high-pressure LC. Nevertheless, the availability of thermally stable columns has steadily improved and new innovations in this area have continued to emerge. This paper gives a brief review and updates on the recent advances in high-temperature liquid chromatography (HTLC). Recent efforts of hyphenating the capillary HTLC with mass spectrometry via a super-atmospheric pressure electrospray ionization is also reported.

Information related to the author
© 2019 Lee Chuin Chen. This is an open access article distributed under the terms of Creative Commons Attribution License, which permits use, distribution, and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.