医用画像情報学会雑誌
Online ISSN : 1880-4977
Print ISSN : 0910-1543
ISSN-L : 0910-1543
CTにおける信号雑音比による低コントラスト分解能の評価
市川 勝弘原 孝則丹羽 伸次山口 功大橋 一也
著者情報
ジャーナル フリー

2007 年 24 巻 3 号 p. 106-111

詳細
抄録

In this paper, we proposed a new quantitative evaluation method of low-contrast detectability for computed tomography(CT)images. It is well known that the contrast-noise-ratio is not enough for CT images to evaluate the low-contrast detectability as image quality index corresponding to various images with different frequency properties. In order to overcome this problem, we adopted a signal-to-noise ratio based on the matched filter model using low-contrast rod images (disc images) as the input signal. The low-contrast detectability indices obtained from this analysis method indicated good agreement with results of perceptual detectability studies. Even for the images obtained by various rod sizes, tube current settings and reconstruction kernels, this method derived appropriate index values matching the results of perceptual studies. The method proposed in this paper is very useful for CT to evaluate quantitative low-contrast image quality.

著者関連情報
© 2007 医用画像情報学会
前の記事 次の記事
feedback
Top