液晶モニタにおけるマンモグラム用ファントムの検出度とSN比の検討
佐藤 真知子犬井 正男茅原 功伊達 隼人中越 智恵子東 吉彦藤澤 健一
著者情報
ジャーナル フリー

2007 年 24 巻 3 号 p. 112-116

詳細
抄録

With the spread of the digital mammography systems, imaging tools for interpreting mammograms are changing from the screen/film to the monitors. In this paper, we assess the effect of the pixel pitch of the Liquid crystal monitors on the detectability of abnormalities. The result of the visual test using mammographic phantoms showed that the 5M-monitor is necessary in the detection of the simulated calcifications, while the 3M-monitor performed equally as the 5M-monitor for detecting the simulated masses and fibrillar structures. The SNR calculated for the phantoms, which incorporate the visual characteristics, well explained the result of the visual test. Thus, it can be an index for evaluating the monitor performance.