Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Major Papers
Detection of Weak Magnetic Fields Induced by Electrical Currents with MRI: Theoretical and Practical Limits of Sensitivity
Tomohisa HATADAMasaki SEKINOShoogo UENO
著者情報
ジャーナル オープンアクセス

2004 年 3 巻 4 号 p. 159-163

詳細
抄録
Detection of weak magnetic fields induced by neuronal electrical activities with magnetic resonance imaging (MRI) is a potentially effective method for functional imaging of the brain. In this study, we compared the theoretical and practical limits of sensitivity for detecting weak magnetic fields with a columnar phantom. The theoretical limit of sensitivity was estimated from signal and noise intensities in magnetic resonance images. The theoretical limit of sensitivity was approximately 10-8T. The practical limit was 10 times the theoretical limit. The dependence of the theoretical limit of sensitivity on acquisition parameters, such as the repetition time (TR), echo time (TE), number of pixels, and spectral width, was quantitatively evaluated. The results indicated the existence of an optimal value in TE/T2*.
著者関連情報
© 2004 by Japanese Society for Magnetic Resonance in Medicine
前の記事 次の記事
feedback