Technical Notes
Parallel Imaging of Head with a Dedicated Multi-coil on a 0.4T Open MRI
Masahiro TAKIZAWATakahide SHIMODAMasayuki NONAKAHiroyuki MOCHIZUKIShinji KAWASAKIHiroyuki TAKEUCHIMiki TACHIBANATetsuhiko TAKAHASHI
Author information
JOURNAL OPEN ACCESS

2005 Volume 4 Issue 2 Pages 95-101

Details
Abstract

Background: Parallel imaging is widely used for cylindrical magnetic resonance imaging (MRI); however, few studies apply parallel imaging to open MRI. We previously developed a parallel method called “RAPID” (rapid acquisition through a parallel imaging design) for imaging the heart on a 0.7T open MRI apparatus, and we have now developed a RAPID head coil and shading correction algorithm for imaging the brain with a 0.4T open MRI apparatus. Images acquired with RAPID were compared with those acquired using a conventional quadrature-detection (QD) head coil.
Materials and Methods: The images were acquired using a dedicated 4-channel RF receiving coil consisting of a solenoid coil and surface coils. For MRI of the brain, we developed 2 methods to acquire the necessary calibration data: a pre-scan method that acquires the calibration data before the main scans and a self-calibration method that acquires the calibration data and imaging data simultaneously. We also modified the algorithm for calculating the shading distribution so that it only uses acquired image data and then corrects the shading.
Results: RAPID was applied for T1-weighted, T2-weighted, fluid-attenuation inversion recovery (FLAIR), time-of-flight (TOF), and diffusion-weighted echo-planar (DW-EPI) imaging. The RAPID images had no visible unfolded artifacts or motion artifacts. Images with the same contrast as that with a conventional QD coil were acquired using the RAPID coil and shading correction.
Conclusion: These preliminary results show that RAPID can be applied to imaging of the head using a 0.4T open MRI apparatus.

Content from these authors
© 2005 by Japanese Society for Magnetic Resonance in Medicine