整形外科と災害外科
Online ISSN : 1349-4333
Print ISSN : 0037-1033
ISSN-L : 0037-1033
脊椎インストゥルメンテーション手術の合併症
中村 孝文池田 天史千田 治道菊池 太朗池辺 顕嗣朗高木 克公
著者情報
ジャーナル フリー

1996 年 45 巻 1 号 p. 33-35

詳細
抄録

We performed a study to evaluate the complication ratio associated with spinal instrumentation surgery for non-deformity or non-tumor patients. One transient cervical myelopathy, one infecion with MRSA and one pedicle fracture were observed. Six cases (8%) had instrument problems, 3 wire breakages and 3 dislodgements of the screw or the rod. It is considered that the most important factor for avoiding complications when performing instrumentation surgery is to select a familiar method. Furthermore, as there was no direct injury to large vessels or nerves by the screw, pedicular screwing was considered to be a safe method.

著者関連情報
© 西日本整形・災害外科学会
前の記事 次の記事
feedback
Top