Advantages and Disadvantages of Multi-axis Intraoperative Angiography Unit for Percutaneous Pedicle Screw Placement in the Lumbar Spine
Hiroto KAGEYAMA, Shinichi YOSHIMURA, Kazutaka UCHIDA, Tomoko IIDA
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: oa.2017-0059

Details
Abstract

We analyzed clinical usefulness of the high resolution imaging system in a hybrid operation room (OR) for posterior lumbar interbody fusion. A total of 17 patients with lumbar spondylolisthesis between February 2014 and August 2016 were included. Multi-axis imaging system in a hybrid OR was used in 12 patients (hybrid OR group); the conventional C-arm fluoroscopy, in 5 patients (C-arm group). The time to confirm the first percutaneous pedicle screw (PPS) angle (hybrid OR, 80 vs C-arm, 249 s; P = 0.0026) and the second to the last PPS angle (77 vs 90 s; P = 0.040) were shorter in the hybrid OR group. Placement accuracy was higher in the hybrid OR group (88.0 vs 59.1%; P = 0.010). Irradiation dose was significantly lower in the C-arm group (462 vs 102 mGy; P = 0.0013). This study suggested that the accuracy of PPS placement and time to confirm the PPS angle are the advantages in a hybrid OR.

Content from these authors
© 2017 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/