遺残クリップのMRI検査における影響―物理的影響と画像上のartifactの検討―
蜂巣 忠山田 英夫後藤 剛貞松下 一之松本 康久楠目 健一北澤 恵美子佐藤 慎一西村 元伸栗山 根廣祖山 暁子藤井 隆之
著者情報
キーワード: 内視鏡的クリップ, MRI
ジャーナル フリー

1997 年 50 巻 p. 226-227

詳細
抄録
The endoscopic clip is made from austenite of stainless steel which consist of Fe, Ni, and Cr. The austenite does not have magnetism its-self, however, slight magnetism is produced in the plastic workings during manufacturing of the clips. And there were no reports with respect to the affect of endoscopic clip on MRI. Therefore, the experimental studies on physical injury and imaging artifact of the endoscopic clip associated with MRI were carried out.
The materials in this studies were 12 endoscopic clips (MD-850) which grasped a chunk of fresh beef, and a MRI equipment used was the Gyroscan, ACS-II (1.5T) . Scanning condition was spine-echo (T1-weighted image, T2-weighted image, proton density-weighted image) and field-echo similar to clinical usage. Total time of the scanning was one hour. All clips remained on the beef throughout the scanning, and there were no physical injuries around them. However, the imaging artifact of the endoscopic clip became bigger in such an order as, T1-weighted image, proton density-weighted image, T2-weighted image and field-echo. We conclude that MRI examination being carried out with the remaining endoscopic clips is a safe procedure.