Reference values for laboratory hematology tests using the automated hematology analyzer system COULTER STKS
Part1. A study on hematological reference intervals for erythrocyte count, hemoglobin value, hematocrit and other erythrocyte parameters (MCV ·MCH ·MCHC and RDW)
YOKO TABENAOTAKE SATOKAZUNORI MIYAKE
Author information
JOURNAL FREE ACCESS

1996 Volume 42 Issue 3 Pages 342-348

Details
Abstract
Hematological reference intervals were established by using COULTER STKS (automatic blood cell counter and cell analyzer based on flow system). For establishment of reference intervals, 529 healthy samples were obtained from 336 males and 193 females between 18 and 67 years. Sex dependent differences in the reference intervals of RBC, Hb, PCV and MCV were calculated. No sex difference was observed in the reference intervals of MCV, MCH and RDW. In the statistical treatment, adjustment by the XP-Type variable transformation was often necessary to obtain the normal distribution. Because of the strong deviation in the distribution pattern, the reference intervals for RDW, MCH, and MCHC were generated by a non-parametric (percentile) method instead of parametric method.