1974 Volume 23 Issue 3 Pages 149-154
Cyclic activation analysis method using 241Am-Be neutron source was studied to determine the fluorine content. In order to effectively accumulate the counting value of the induced nuclide, the time for neutron burst, for counting and for sample transport considering the half life of induced nuclide were discussed. See saw type and pneumatic tube type cyclic activation analysis systems were deviced to accumulate the activity of 16N produced by (n, α) reaction. Applying the result, the fluorine content of several kinds of fluorine plastics containing various kind of fillers was determined. The result obtained by cyclic activation method agreed with that by chemical analysis.