改良型ピンホールコリメータの性能試験と小動物応用
高野 英明郭 東英佐野 友昭前田 静一後藤 邦彦金子 昌幸
著者情報
ジャーナル フリー

1988 年 37 巻 8 号 p. 456-459

詳細
抄録

Testing performance of the improved pinhole collimator of 1.5 mm aperture attached to the gamma camera was carried out with three kinds of radionuclides, 99mTc, 67Gaand131I, using a hand-made phantom. The results obtained in this experiment were compared to those obtained with a human pinhole collimator of 4 mm aperture. Those were nearly understood by taking account of the effective aperture of a pinhole collimator for each effective photon energy. And all scintigrams obtained from rats with various scintigraphy showed high resolution images.

著者関連情報
© 社団法人 日本アイソトープ協会