133Xeガスを用いた肺換気シンチグラフィ検査における術者の被曝線量の推定
辻 成人中司 等山本 吉之阿知波 一生柴田 靖彦春藤 英雄竹内 憲彦住本 吾一岩田 宏道
著者情報
ジャーナル フリー

1991 年 40 巻 4 号 p. 153-156

詳細
抄録

133Xe Gas concentration during the examinations of the lung ventilating scintigraphy using 133Xe for three cases, was measured continuously by the flow type ionization chamber to estimate effective dose equivalent to the operator. The air flow distribution of the examination room was also measured for reference to examining the behavior of 133Xe gas. The air around the patient lingers for few minutes or more. Based on the derived air concentration (DAC) in ICRP Publication 30, effective dose equivalent to the operator per examination was estimated to be 0.1-1.0μSv, which was lower than expected.