RADIOISOTOPES
Online ISSN : 1884-4111
Print ISSN : 0033-8303
ISSN-L : 0033-8303
201Tl心筋SPECTにおけるcircumferential profile解析法の検討
―異なる代表値による比較―
清水 光春柳 英清中川 富夫佐藤 修平新屋 晴孝竹田 芳弘平木 祥夫永谷 伊佐雄
著者情報
ジャーナル フリー

1997 年 46 巻 5 号 p. 284-290

詳細
抄録

Circumferential profile analysis is one of quantitative analyses in 201Tl myocardial scintigraphy. We compared the detectability of coronary artery stenosis between 4 various methods of circumferential profiles (peak point method, all points method, 3 points around the peak method and average method) based on short-axial images of 201Tl myocardial single photon emission computed tomography. Diagnostic criteria for coronary artery stenosis was that the circumferential profile curve of a patient was below the normal limit (mean—2SD) obtained from 23 normal volunteers in more than serial 3 profiles. After examining 16 patients with coronary artery stenosis demonstrated by coronary arteriography, it was concluded that both peak point method and 3 points around the peak method surpassed other methods in the detectability of clinically significant coronary artery stenosis.

著者関連情報
© 社団法人 日本アイソトープ協会
前の記事 次の記事
feedback
Top