日本レオロジー学会誌
Online ISSN : 2186-4586
Print ISSN : 0387-1533
ISSN-L : 0387-1533
ノート
汎用水晶振動子マイクロバランスを用いた10MHz域の動的粘性率評価
吉澤 僚太浦川 理井上 正志
著者情報
ジャーナル フリー

2022 年 50 巻 4 号 p. 333-337

詳細
抄録

This study examined the validity of viscosity measurements in the frequency region of 10 MHz using a commercially available quartz crystal microbalance (QCM). Ethylene glycol aqueous solutions were used as model samples. From the changes in the resonant frequency and dissipation parameter before and after the introduction of the sample liquid, we determined dynamic viscosity through Johannsmann's equation and compared it with literature data for zero shear viscosity. As a result, the viscosity estimated by QCM was systematically higher. Based on the analysis of the electrical equivalent circuit of the resonator, we considered that such deviation is due to the energy loss inside the quartz resonator and the non-ideal nature of the interface in contact with the sample liquid. To avoid such non-idealities, we proposed a method of applying a calibration equation to the viscosity value obtained from the Johannsmann equation. This makes it possible to easily measure viscosity in the 10 MHz regions with an accuracy of ± 3 % error.

著者関連情報
© 2022 一般社団法人 日本レオロジー学会
前の記事 次の記事
feedback