Study of the Immersion Characteristic of Thermocouples with a Heat-pipe up to 960-degree
Masahiko Gotoh
著者情報
会議録・要旨集 フリー

p. 111

詳細
抄録
In this paper results of the measurements on the irregular thermal emf by means of the thermoelectric scanning are reported. Scanning is carried out in the pressure controlled heat pipe at 960°C over 50cm along the thermocouples. Measurement was performed with an Au/Pt, two Pt/Pd and a multi element wire thermocouples. One of the Pt/Pd thermocouples was subjected to 500 hours exposure to 1300°C air environment. The other Pt/Pd thermocouple is newly fabricated one and act as the reference. There found no significant difference in the spurious Emf between the two Pt/Pd thermocouples. Discrepancy due to the different immersion was 55mK for the two Pt/Pd at most. This result already suggests that they are superior to the type S and R thermocouples in terms of the immersion characteristics. It is very important to access the source of uncertainty of thermocouples due to the different immersion profile of the furnace. The study provides more realistic information for this purpose.