Improvement In Temperature Scale Realization Above The Copper Point
Naoko YamamotoYoshiro Yamada
著者情報
会議録・要旨集 フリー

p. 9

詳細
抄録
For improvement of the high-temperature scale realization, the spectral responsivity measurement uncertainty, which is the dominant source of uncertainty in the high temperature range, was investigated. A new single grating monochromator was introduced for this investigation, by means of which the target uncertainty of 0.03 nm was realized.