Factors determining maximum torque and achievement of the recommended torque for manual implant drivers: A pilot study
Tomoki NishiuchiYuji SatoNoboru KitagawaTokiko OsawaAkio IsobeHiroki Shiba
著者情報
キーワード: manual driver, maximum torque, implant
ジャーナル フリー

2021 年 33 巻 3 号 p. 82-87

詳細
抄録

When fixing an oral implant superstructure with a screw, operators must be aware of the torque being applied by their fingers to prevent the transmission of excessive or insufficient torque to the implant. In this study, we identified the factors that determine individual maximum attainable torque and those that determine the achievement of the prescribed torque. We evaluated 16 dentists on their use of two types of manual implant drivers(UniGrip by Nobel Biocare and Carrier Hex by Zimmer Biomet)and measured the maximum torque(MT)generated by their fingers. The target torque was set at 15N. Measurements were taken while the participants were turning the implant screw with or without gloves in both clockwise and counterclockwise directions. The grip and finger strength of each participant were measured, and the data showed that torque values were higher among the male participants during clockwise rotation and when they were wearing gloves(p<0.05). Positive correlations were found between the MT and grip strength and between the MT and finger strength. These results suggest that dentists should monitor their ability to consistently achieve the recommended torque for implant drivers.