Tetsu-to-Hagane
Online ISSN : 1883-2954
Print ISSN : 0021-1575
ISSN-L : 0021-1575
Regular Article
Determination of Oxygen in Magnesium by Glow Discharge Mass Spectrometry
Emi KIKUTAHitoshi ASANOTadashi KIKUCHI
Author information
JOURNAL OPEN ACCESS

2007 Volume 93 Issue 2 Pages 128-131

Details
Abstract
In the present experiments, preferable results were obtained by applying glow discharge mass spectrometry (GD-MS) for quantitative analysis of oxygen in magnesium. The relative sensitivity factor (RSF) method was applied for measurements, since there was no standard material for magnesium analysis. Removal of the surface oxygen contaminants was investigated before the Mg specimen measurements. It was clarified that 5000 s of pre-discharge period was required to remove surface oxygen for both of pin shape and disk shape samples. As a result, the oxygen contents of the pure Mg and re-dissolved Mg were obtained as 9 ppm and 26 ppm, respectively. In order to check the results obtained by GD-MS measurements, charged particle activation analysis was also applied for the same specimens. The oxygen contents of the pure Mg and re-dissolved Mg were obtained as 12 ppm and 26 ppm, respectively with this method. On the basis of the results, GD-MS is possible technique for oxygen analysis in Mg and its alloys with high sensitivity.
Content from these authors
© 2007 The Iron and Steel Institute of Japan

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback