The Tohoku Journal of Experimental Medicine
Online ISSN : 1349-3329
Print ISSN : 0040-8727
ISSN-L : 0040-8727
Development of an Apparatus to Measure Plasma Colloid Osmotic Pressure Mounted on the Extracorporeal Hemodialysis Circuit
RYOETSU ABESHIGEMI OOISHIKAZUTO SATOHIROSHI SUMATSUNEMICHI WAKABAYASHISATOSHI SAWATANIHIROYUKI TAKEUCHI
Author information
JOURNAL FREE ACCESS

1995 Volume 176 Issue 3 Pages 149-154

Details
Abstract
ABE, R., OOISHI, S., SATO, K., SUMA, H., WAKABAYASHI, T., SAWATANI, S. and TAKEUCHI, H. Development of an Apparatus to Measure Plasma Colloid Osmotic Pressure Mounted on the Extracorporeal Hemodialysis Circuit. Tohoku J. Exp. Med., 1995, 176 (3), 149-154-We newly developed an apparatus to measure colloid osmotic pressure (COP) of blood plasma continuously by hemodialysis (HD) or extracorporeal ultrafiltration method (ECUM) without any invasive procedure or blood sampling. The COP value measured by this apparatus highly correlated with the COP value calculated from the concentration of total protein of blood plasma (TP) sampled simultaneously (y=21.182+0.96256x, r2=0.911). Measurement of the COP with this apparatus seems to be reliable.
Content from these authors
© Tohoku University Medical Press
Previous article Next article
feedback
Top