Studies on the Diffractometric Measurement of the Diameter of Erythrocytes
II. Accuracy of the diffractometric method of Haden-Hausser's erythrocytometer in cases of blood with pathological anisocytosis
松田 幸次郎
著者情報
ジャーナル フリー

1941 年 40 巻 3-4 号 p. 330-335

詳細
抄録

(1) Accuracy of Haden-Hausser's erythrocytometer was tested in cases of pathologically marked anisocytosis with 25 blood samples.
(2) The result showed an error up to 0.5μ or more may occur in such cases, especially if the inner red ring is used as was suggested by Haden.
(3) After all, the yellow ring is the best for the examination of any sort of blood samples.