Annual Meeting of the Japanese Society of Toxicology
The 48th Annual Meeting of the Japanese Society of Toxicology
Session ID : P-145E
Conference information

e-Poster
Usefulness of the pulse oximeter method for the evaluation of heart rate in conscious rats
*Daigo IMOTOKenta WATANABETadashi TSUBOUCHIKazuhiro CHIHARAIzuru MIYAWAKI
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Details
Abstract

Pulse oximeters can be a powerful tool to measure heart rate (HR) noninvasively, but their accuracy is still unclear. Therefore, we investigated the accuracy by comparing HR between the pulse oximeter (MouseOx Plus®) and the telemetry method in telemetered conscious rats administered vehicle, milrinone, or atenolol. As a result, there was a good correlation in the HR with the range from 240 to 435 bpm (R2=0.9996). The significant changes in HR noted in the telemetry method were well noted also in the pulse oximeter. The pulse oximeter can be brief but accurate method for the evaluation of HR.

Content from these authors
© 2021 The Japanese Society of Toxicology
Previous article Next article
feedback
Top