2012 Volume E95.D Issue 3 Pages 829-834
We have developed a portable NIRS-based optical BCI system that features a non-invasive, facile probe attachment and does not require muscle movement to control the target devices. The system consists of a 2-channel probe, a signal-processing unit, and an infrared-emission device, which measures the blood volume change in the participant's prefrontal cortex in a real time. We use the threshold logic as a switching technology, which transmits a control signal to a target device when the electrical waveforms exceed the pre-defined threshold. Eight healthy volunteers participated in the experiments and they could change the television channel or control the movement of a toy robot with average switching times of 11.5±5.3s and the hit rate was 83.3%. These trials suggest that this system provides a novel communication aid for people with motor disabilities.