日本原子力学会 年会・大会予稿集
2012 Annual Meeting
セッションID: O35
会議情報
半導体検出器・低温動作型検出器
Development of superconducting TES microcalorimeters for hard X-ray and gamma-ray spectroscopy
*Thushara Damayanthi Rathnayaka MudiyanselageMasashi OhnoSuiichi HatakeyamaHiroyuki TakahashiKoji TakasakiKeisuke Maehata
著者情報
会議録・要旨集 フリー

詳細
抄録
We are developing bulk absorber coupled superconducting TES microcalorimeters for high energy resolution gamma-ray spectroscopy. Our first prototype Sn absorber epoxy-post coupled TES detector was measured at JAEA Oarai research and development center, reported ~ 485 eV FWHM energy resolution at 60 keV, with a fall time of the signal response ~6.6 ms.
著者関連情報
© 2012 Atomic Energy Society of Japan