1992 年 25 巻 1-2 号 p. 65-69
To measure cytochrome P-450 (P-450) content in tissue sections, a dual wavelength microphotometry system linked with a personal computer was developed. A computer program to calculate P-450 content in sections was also established. With this system, accuracy and microphotometric efficiency of the measurement of P-450 content in sections were improved markedly. Furthermore, this system can analyze many data derived from numerous microphotometric spots in sections using the program during measurement. Therefore, the measurement of P-450 content in tissue sections can be done precisely and rapidly with this system.