Determination of Lead in Whole Blood by Filter Furnace Laser-Excited Atomic Fluorescence Spectrometry
Ricardo Q. AUCÉLIOBenjamin W. SMITHJames D. WINEFORDNER
著者情報
ジャーナル フリー

1999 年 15 巻 4 号 p. 321-325

詳細
抄録
Filter furnace laser-excited atomic fluorescence spectroscopy was used for the determination of lead in whole blood. The use of a Katskov filter has enabled the direct comparison between lead fluorescence signals obtained with blood standards and water standards. Only a simple dilution of the blood was required. This direct correlation has allowed the interpolation of the lead signal of blood samples using a lead water standard calibration curve. NIST certified blood samples were analyzed using the method with excellent results.