Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Lead-Selective Membrane Potentiometric Sensor Based on an 18-Membered Thiacrown Derivative
Mojtaba SHAMSIPURMohammad Reaza GANJALIAhmad ROUHOLLAHI
著者情報
ジャーナル フリー

2001 年 17 巻 8 号 p. 935-938

詳細
抄録

A PVC-based membrane electrode for lead ions based on hexathia-18-crown-6-tetraone as membrane carrier was prepared. The influence of membrane composition, pH of test solution and foreign ions on the electrode performance were investigated. The electrode showed a Nernstian response over a lead concentration range from 1.0 × 10-6 to 8.0 × 10-3 M at 25°C, and was found to be very selective, precise and usable within the pH range 3.0 - 6.0. The electrode was successfully used as an indicator electrode in potentiometric titration of lead ions and in direct determination of lead in water samples.

著者関連情報
© 2001 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top