Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Clotrimazole-Triiodide Ion Association as an Ion Exchanger for a Triiodide Ion-Selective Electrode
Khalil FARHADIRamin MALEKI
著者情報
ジャーナル フリー

2002 年 18 巻 2 号 p. 133-136

詳細
抄録
A novel triiodide ion-selective electrode based on a clotrimazole-triiodide ion pair as a membrane carrier was prepared. It has a linear response to triiodide from 8 × 10-6 to 5 × 10-3 M with a slope of -68.9 mV per decade and a detection limit of 5 × 10-6 M. The electrode response is independent of the pH of the solution in the pH range 2 - 9. It has a very short response time and can be used for at least 3 months without any considerable divergence in the potentials. The proposed sensor revealed very good selectivities for I3- over a variety of other anions. It was used as an indicator electrode in the potentiometric titration of triiodide ions and in an indirect potentiometric determination of clotrimazole in pharmaceutical preparations.
著者関連情報
© 2002 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback