Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Determination of Trace Amounts of Thallium by Adsorptive Cathodic Stripping Voltammetry with Xylenol Orange
Esmaeil SHAMSMehdi YEKEHTAZ
著者情報
ジャーナル フリー

2002 年 18 巻 9 号 p. 993-996

詳細
抄録
Trace amounts of thallium(I) can be determined using adsorptive cathodic stripping voltammetry in the presence of Xylenol Orange (XO). The reduction current of the thallium(I)-XO complex ion was measured by square-wave cathodic stripping voltammetry. The peak potential was at -0.44 V vs. Ag/AgCl. The effect of various parameters (pH, ligand concentration, accumulation potential and collection time) on the response are discussed. The response was linearly related to the thallium concentration in the range 0.5 - 110 ng ml-1 and 110 - 2000 ng ml-1. The limit of detection was 0.2 ng ml-1. The relative standard deviation for the determination of 80 ng ml-1 thallium was 2.8%. Many common anions and cations did not interfere with the determination of thallium. The interference of lead was reduced by the addition of 0.003 M sodium carbonate. The voltammetric procedure was then successfully applied to the determination of thallium in various complex samples.
著者関連情報
© 2002 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top