Non-destructive Rapid Analysis of Brominated Flame Retardants in Electrical and Electronic Equipment Using Raman Spectroscopy
Sonoko KIKUCHIKazuteru KAWAUCHISadatsugu OOKIMasaru KUROSAWAHisashi HONJHOTeruo YAGISHITA
著者情報
ジャーナル フリー

2004 年 20 巻 8 号 p. 1111-1112

詳細
抄録
Using Raman spectroscopy without any preparation, we analyzed the brominated flame retardant, polybromodiphenyl ethers (PBDEs), which has been prohibited by the European Union. The typical Raman bands of PBDEs, which are clearly different from those of other brominated flame retardants, are from 100 to 300 cm-1. In our analysis, the detection limit was about 100 ppm, and the analysis took about 1 min, whereas the usual method of solvent extraction using gas chromatography/mass spectroscopy takes 50 h.
著者関連情報
© 2004 by The Japan Society for Analytical Chemistry