Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Spectrophotometric Determination of Urinary Protein with o-Sulfophenylfluorone-Metal Complex
Takako YAMAGUCHIEmi AMANOShinichiro KAMINOSatoko UMEHARAChizuko YANAIHARAYoshikazu FUJITA
著者情報
ジャーナル フリー

2005 年 21 巻 10 号 p. 1237-1240

詳細
抄録
A simple and sensitive spectrophotometric method for the determination of human serum albumin (HSA) was established based on the ternary complex-formation reaction of HSA with o-sulfophenylfluorone (SPF) as a xanthene dye and metal ion (niobium(V) and bismuth(III)) in the presence of a dispersion agent. This new method enabled the determination of HSA in the range of 1 - 15 µg/ml HSA by measuring the difference of the absorbance at 530 nm between HSA-SPF-metal ion and SPF-metal ion solutions. In the determination of HSA, this method is about 2-times more sensitive than the Pyrogallol Red-molybdenum(VI) method (PR method), which accounts for more than 80% of the quantification methods for urinary protein assays in Japan. There was no significant difference between the results obtained by the present method and the PR method for human urine samples. The binding process between the SPF-metal complex and HSA was studied by determining the binding parameters and the thermodynamic parameters.
著者関連情報
© 2005 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top