Determination of Wear Metals in Lubrication Oils: A Comparison Study of ICP-OES and FAAS
Pekka VÄHÄOJA, Ilkka VÄLIMÄKI, Katri HEINO, Paavo PERÄMÄKI, Toivo KUOKKANEN
著者情報
ジャーナル フリー

2005 年 21 巻 11 号 p. 1365-1369

詳細
抄録
Certain wear metals (Fe, Cr, Ni, Cu and Zn) of various lubrication oils were determined by means of ICP-OES and FAAS. The kerosene dilution method, which is used widely together with ICP-OES, was applied with both methods here. Calibration standards were made from a commercial organo-metallic standard. Our aim was to clarify the possibility of using the quick kerosene dilution method together with FAAS for a rapid check for certain indicator metals. Metal determinations with FAAS were accurate enough for quantitative work in machine condition diagnostics and waste oil characterization, when compared with those with ICP-OES.
著者関連情報
© 2005 by The Japan Society for Analytical Chemistry