Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Indirect Determination of Uranium by the On-line Reduction and Fluorimetric Detection of Cerium(III)
Chatchalida BOONPANAIDKoichi OGUMA
著者情報
ジャーナル フリー

2005 年 21 巻 2 号 p. 155-159

詳細
抄録
A novel flow injection method has been developed for the indirect determination of uranium by the on-line reduction and subsequent fluorimetric detection of cerium(III). A sample solution containing uranium(VI), prepared as a sulfuric acid solution, was injected into a sulfuric acid carrier solution and passed through a column packed with metal bismuth to reduce uranium(VI) to uranium(IV). The sample solution was merged with a cerium(IV) solution to oxidize uranium(IV) to uranium(VI) and the cerium(III) generated was then monitored fluorimetricaly. The present method is free from interference from zirconium, lanthanides, and thorium, and has been successfully applied to the determination of uranium in monazite coupled with an anion-exchange separation in a sulfuric acid medium to eliminate iron(III). The sample throughput was 25 per hour and the lowest detectable concentration was 0.0042 mg l-1.
著者関連情報
© 2005 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top