Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
An Amperometric Glucose Biosensor Based on Glucose Oxidase Immobilized in Electropolymerized Poly(o-aminophenol) and Carbon Nanotubes Composite Film on a Gold Electrode
Dawei PANJinhua CHENShouzhuo YAOWenyan TAOLihua NIE
著者情報
ジャーナル フリー

2005 年 21 巻 4 号 p. 367-371

詳細
抄録
An amperometric glucose biosensor is developed that is based on immobilization of glucose oxidase (GOD) in a composite film of poly(o-aminophenol) (POAP) and carbon nanotubes (CNT), which are electrochemically co-polymerized at a gold (Au) electrode. Because of the high surface per volume ratio and excellent electrical conductivity of CNT, the biosensor based on an Au/POAP/CNT/GOD electrode has lower detection limit (0.01 mM), larger maximum response current (0.24 mA cm-2) and higher sensitivity (11.4 mA M-1 cm-2) than the values of the biosensor based on an Au/POAP/GOD electrode. Additionally, the biosensor shows fast response time, large response current, and good anti-interferent ability for ascorbic acid, uric acid and acetaminophen. Good reproducibility and stability of the biosensor are also observed.
著者関連情報
© 2005 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top