Atropine-selective Membrane Electrodes and Relative Selectivity Concept
Mohsen M. ZAREH
著者情報
ジャーナル フリー

2008 年 24 巻 7 号 p. 889-894

詳細
抄録
TFPB was introduced as a charged ionophore for atropine selective electrodes. Typical Nernstian responses were found (57.78, 58.95 and 58.41 mV/decade) for PVC-membrane electrodes incorporating NPOE, DOS, and DDP as plasticizers. They exhibited practical linear ranges of 9.1 × 10-3 - 10-6, 9.1 × 10-3 - 10-6 and 9.1 × 10-3 - 10-7 M, respectively. It works in the sub-micro scale of atropine concentrations. The optimum pH-range was 3.18 - 8.97. The selectivity coefficient values were estimated for different organic and inorganic cations. They were interpreted by using the “Relative Selectivity Concept”, which was introduced for the first time. The new concept was applied for comparing the selectivity properties of previously reported electrodes. The effect of the presence of ephedrine, caffeine, glucose, Na+, Ca2+, and Mg2+ on the calibration graphs of the electrodes was studied.