Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Online Concentration by Head-column Field-amplified with Large-volume Sample Stacking Using Flow Injection-Capillary Electrophoresis for the Analysis of Four Active Components in Cold Medicines
Xiumei LIUJingshu ZHANGXingguo CHEN
Author information
JOURNAL FREE ACCESS

2008 Volume 24 Issue 8 Pages 1031-1037

Details
Abstract

A simple, effective, and sensitive online concentration method for the detection of dextromethorphan hydrobromide (Dex), chlorphenamine hydrogen maleate (Chl), pseudoephedrine hydrochloride (Pse) and paracetamol (Par) based on flow injection-capillary electrophoresis (FI-CE) analysis with head-column field-amplified sample stacking and large-volume sample stacking was developed. The background electrolyte (BGE) was a solution composed of 55 mM borate-15% (v/v) acetonitrile (ACN) (pH 9.3). The sample was injected electrokinetically between plugs of water. Under the optimum conditions, about a 30-fold improvement in the concentration sensitivity relative to normal CE methods was achieved, giving low limits of detection (LOD) of 1.94 × 10-5, 0.64 × 10-5, 1.16 × 10-5 and 2.84 × 10-5 mg/mL for Dex, Chl, Pse and Par, respectively. The repeatability (defined as RSD) was 1.01, 1.91, 0.89 and 0.92% with the peak-area evaluation and 1.94, 3.98, 2.66 and 3.27% with the peak-height evaluation for Dex, Chl, Pse and Par, respectively. This method has been successfully applied to the analysis of commercial pharmaceutical preparations containing Dex, Chl, Pse and Par.

Content from these authors
© 2008 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top